DocumentCode
273193
Title
A wideband spectrometer for the measurement of permittivity
Author
Pring, P.C.J. ; Smith, M.J.A.
Author_Institution
Warwick Univ., Coventry, UK
fYear
1988
fDate
27-30 Jun 1988
Firstpage
326
Lastpage
327
Abstract
The authors are concerned specifically with the methodology and implementation of an instrument able to measure the complex permittivity of solids over 13 decades of frequency (10-5 Hz to 105 Hz) at temperatures between -185°C and 400°C. They conclude that the optimum methodology at the present time for a sub-gigahertz spectrometer is the balanced capacitor bridge. Operation is necessarily in the frequency domain
Keywords
bridge instruments; permittivity measurement; spectrometers; -185 to 400 C; 1E-5 to 1E5 Hz; balanced capacitor bridge; complex permittivity; measurement of permittivity; permittivity spectrometer; sub-gigahertz spectrometer; wideband spectrometer;