DocumentCode
2732739
Title
Proposition of tolerance requirements adapted for the calibration of a reverberation chamber
Author
Lemoine, Christophe ; Besnier, Philippe ; Drissi, M. Hamed
Author_Institution
IETR, INSA of Rennes, Rennes
fYear
2009
fDate
12-16 Jan. 2009
Firstpage
41
Lastpage
44
Abstract
This paper aims to propose an evolution of tolerance requirements for calibrating a reverberation chamber (RC). Instead of the normative threshold of 3 dB, even more for lower frequencies, we propose a confidence interval based on distribution functions of electric field measurements. Moreover, this involves to take into account the number of independent stirrer positions and the number of probe locations, which are defined in the measurement setup. The proposed requirements are based on measurements. We increase the complexity of the study in a progressive manner in order to propose a more realistic confidence interval.
Keywords
electric field measurement; reverberation chambers; confidence interval; electric field measurements; independent stirrer positions; probe locations; reverberation chambers; tolerance requirements; Calibration; Distribution functions; Electromagnetic compatibility; Electromagnetic measurements; Erbium; Frequency estimation; Frequency measurement; Position measurement; Probes; Reverberation chamber;
fLanguage
English
Publisher
ieee
Conference_Titel
Electromagnetic Compatibility, 2009 20th International Zurich Symposium on
Conference_Location
Zurich
Print_ISBN
978-3-9523286-4-4
Type
conf
DOI
10.1109/EMCZUR.2009.4783385
Filename
4783385
Link To Document