Abstract
Planar 2-dimensional dielectrophoresis electrode geometries are limited in only being capable of handling fluid volumes ranging from picolitres to hundreds of microliters per hour. A 3-dimensional electrode system has been developed capable of handling significantly larger volumes of fluid. Using finite element modeling the electric field distribution within various bore sizes was realized. From these simulations it is possible to optimize bioprocessing factors influencing the performance of a dielectrophoretic separator. Process calculations have shown that flow-rates of 25ml hr-1 or more can be attained for the separation of heterogeneous populations of bio-particles based on their dielectric properties.
Keywords
biological techniques; cellular biophysics; electrophoresis; finite element analysis; flow separation; 3 dimensional dielectrophoretic flow separator device; bio-particles; dielectric property; electric field distribution; finite element modelling; optimal bioprocessing condition; Biomedical electrodes; Boring; Cells (biology); Dielectrics; Dielectrophoresis; Finite element methods; Geometry; Insulation; Particle separators; Throughput; 3-D modeling; Dielectrophoresis; bioprocess;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 2004. IEMBS '04. 26th Annual International Conference of the IEEE
Conference_Location
San Francisco, CA
Print_ISBN
0-7803-8439-3
Type
conf
DOI
10.1109/IEMBS.2004.1403744
Filename
1403744
Link To Document