DocumentCode
2738959
Title
End tidal carbon dioxide measurement using an electro acoustic sensor
Author
Folke, M. ; Hök, B. ; Ekström, M. ; Bäcklund, Y.
Author_Institution
Department of Electrical Engineering, Mälardalen University, Västerås, Sweden
Volume
2
fYear
2004
fDate
1-5 Sept. 2004
Firstpage
3452
Lastpage
3455
Abstract
End tidal carbon dioxide measurement with an electro-acoustic sensor is demonstrated. The sensor consists of an acoustic resonator coupled to a low cost electro-acoustic element. By simultaneous measurements with a reference sensor, the new device was tested on subjects performing exercise, hypo- and hyperventilation whereby the CO2 concentration ranged from 2.1 to 7.0 kPa. The output from the experimental device correlated well with the reference CO2 readings with a correlation coefficient of 0.976. Response time for expiration less than 0.8 seconds was noted. The new device could be useful in situations where selectivity to other gases is not important.
Keywords
End tidal; capnometry; carbon dioxide; electro acoustic sensor; Acoustic devices; Acoustic measurements; Acoustic sensors; Acoustic testing; Carbon dioxide; Chemical elements; Costs; Delay; Gas detectors; Performance evaluation;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 2004. IEMBS '04. 26th Annual International Conference of the IEEE
Print_ISBN
0-7803-8439-3
Type
conf
DOI
10.1109/IEMBS.2004.1403969
Filename
1403969
Link To Document