• DocumentCode
    2738959
  • Title

    End tidal carbon dioxide measurement using an electro acoustic sensor

  • Author

    Folke, M. ; Hök, B. ; Ekström, M. ; Bäcklund, Y.

  • Author_Institution
    Department of Electrical Engineering, Mälardalen University, Västerås, Sweden
  • Volume
    2
  • fYear
    2004
  • fDate
    1-5 Sept. 2004
  • Firstpage
    3452
  • Lastpage
    3455
  • Abstract
    End tidal carbon dioxide measurement with an electro-acoustic sensor is demonstrated. The sensor consists of an acoustic resonator coupled to a low cost electro-acoustic element. By simultaneous measurements with a reference sensor, the new device was tested on subjects performing exercise, hypo- and hyperventilation whereby the CO2concentration ranged from 2.1 to 7.0 kPa. The output from the experimental device correlated well with the reference CO2readings with a correlation coefficient of 0.976. Response time for expiration less than 0.8 seconds was noted. The new device could be useful in situations where selectivity to other gases is not important.
  • Keywords
    End tidal; capnometry; carbon dioxide; electro acoustic sensor; Acoustic devices; Acoustic measurements; Acoustic sensors; Acoustic testing; Carbon dioxide; Chemical elements; Costs; Delay; Gas detectors; Performance evaluation;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2004. IEMBS '04. 26th Annual International Conference of the IEEE
  • Print_ISBN
    0-7803-8439-3
  • Type

    conf

  • DOI
    10.1109/IEMBS.2004.1403969
  • Filename
    1403969