Title :
A test microchip for evaluation of hermetic packaging technology for biomedical prosthetic implants
Author :
Kazemi, Mohammad ; Basham, Eric ; Sivaprakasam, Mohanasankar ; Wang, Guoxing ; Rodger, Damien ; Weiland, James ; Tai, Y.C. ; Liu, Wentai ; Humayun, Mark
Author_Institution :
Dept. of Biomedical Eng., Southern California Univ., Los Angeles, CA, USA
Abstract :
The development of a test chip that will be used to evaluate a hermetic and biocompatible package for the driving CMOS circuitry of a retinal prosthesis is described. The package design is estimated to be about 2 × 2 × 0.3 mm3 and will be formed by conformal layers of parylene and a metal (e.g. titanium) as inner and outer protections, respectively. The test chip has been specifically designed for evaluation of the packaging technology. It consists of many blocks of analog and digital components as well as relative humidity and temperature sensors. The test chip has more probe points than a typical chip, allowing a more thorough evaluation of circuit behavior during the testing. This chip will first be coated in a layer of parylene C and soaked in heated isotonic saline for an extended period of time. Every block in the chip will then be tested for functionality using the surface probe points. The next step is to coat the surface of another test chip with parylene and a metal and repeat these soak tests. The results will then be analyzed and mean time-to-failure for the different samples will then be computed. Using the accelerated testing paradigm, these results will then be extrapolated to mean time-to-failure in the operating intraocular environment. Parylene test structures have already undergone an accelerated lifetime test and results have been analyzed.
Keywords :
CMOS analogue integrated circuits; CMOS digital integrated circuits; chip scale packaging; eye; humidity sensors; polymer films; prosthetics; temperature sensors; titanium; 0.3 mm; 2 mm; CMOS circuitry; Ti; analog components; biocompatible package; biomedical prosthetic implants; digital components; heated isotonic saline; hermetic packaging technology; intraocular environment; microchip; parylene C; relative humidity; retinal prosthesis; temperature sensors; CMOS technology; Circuit testing; Implants; Life estimation; Packaging; Probes; Prosthetics; Protection; Retina; Titanium; Hermetic packaging; biomedical implants; soak testing; test chip;
Conference_Titel :
Engineering in Medicine and Biology Society, 2004. IEMBS '04. 26th Annual International Conference of the IEEE
Conference_Location :
San Francisco, CA
Print_ISBN :
0-7803-8439-3
DOI :
10.1109/IEMBS.2004.1404142