• DocumentCode
    2743956
  • Title

    Preliminary accuracy investigation of an AC voltage standard based on square wave biased Josephson voltage reference

  • Author

    Nissilä, J. ; Ojasalo, K. ; Maisi, V. ; Kaasalainen, J. ; Behr, R. ; Manninen, A.

  • Author_Institution
    Centre for Metrol. & Accreditation (MIKES), Espoo, Finland
  • fYear
    2012
  • fDate
    1-6 July 2012
  • Firstpage
    58
  • Lastpage
    59
  • Abstract
    This paper describes the first measurement results with a square-wave based Josephson ac voltage standard. Using a planar multijunction thermal converter as a reference the accuracy was found to be in the μV/V range at 1 V level at 62.5 Hz and 1 kHz. A systematic error of 1.3 μV/V was observed.
  • Keywords
    measurement standards; power convertors; reference circuits; voltage measurement; AC voltage standard; frequency 1 kHz; frequency 62.5 Hz; planar multijunction thermal converter; square wave biased Josephson voltage reference; systematic error; voltage 1 V; Accuracy; Arrays; Frequency measurement; Standards; Synthesizers; Transient analysis; Voltage measurement; AC voltage standard; Josephson array; programmable Josephson array; thermal converter;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements (CPEM), 2012 Conference on
  • Conference_Location
    Washington, DC
  • ISSN
    0589-1485
  • Print_ISBN
    978-1-4673-0439-9
  • Type

    conf

  • DOI
    10.1109/CPEM.2012.6250659
  • Filename
    6250659