This paper describes the first measurement results with a square-wave based Josephson ac voltage standard. Using a planar multijunction thermal converter as a reference the accuracy was found to be in the μV/V range at 1 V level at 62.5 Hz and 1 kHz. A systematic error of 1.3 μV/V was observed.
Keywords :
measurement standards; power convertors; reference circuits; voltage measurement; AC voltage standard; frequency 1 kHz; frequency 62.5 Hz; planar multijunction thermal converter; square wave biased Josephson voltage reference; systematic error; voltage 1 V; Accuracy; Arrays; Frequency measurement; Standards; Synthesizers; Transient analysis; Voltage measurement; AC voltage standard; Josephson array; programmable Josephson array; thermal converter;