• DocumentCode
    2745063
  • Title

    Calibration of AC current from 10 µA to 1 mA AT 10 kHz

  • Author

    Liu Yue ; Zhang Xiu-Zeng ; Zhao Hong-Gang ; Zheng Lian-Ying ; Deng Lin

  • Author_Institution
    Div. of Electr. & Magnetics Metrol., Nat. Inst. of Metrol., Beijing, China
  • fYear
    2012
  • fDate
    1-6 July 2012
  • Firstpage
    166
  • Lastpage
    167
  • Abstract
    New AC current calibration equipment is developed, including high output impedance AC current sources, sampling resistors and AC voltage buffers, which is designed for calibrating AC current from 10μA to 1mA, at up to 10kHz. The output impedance of the AC current source is adequate enough to adapt to most of the resistance load. The AC current source is connected to the sampling resistor with a very short wire, and is connected to UUT with a very short wire as well. The calibration uncertainty can be estimated as 68ppm, k=2.
  • Keywords
    calibration; electric current measurement; measurement uncertainty; resistors; wires (electric); AC current calibration equipment; AC voltage buffer; UUT; calibration uncertainty; current 10 muA to 1 mA; frequency 10 kHz; high output impedance AC current source; resistance load; sampling resistor; very short wire; Calibration; Impedance; Metrology; Operational amplifiers; Resistors; Uncertainty; Voltage measurement; AC current; AC voltage buffer; load effect; output impedance; sampling resistor;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements (CPEM), 2012 Conference on
  • Conference_Location
    Washington, DC
  • ISSN
    0589-1485
  • Print_ISBN
    978-1-4673-0439-9
  • Type

    conf

  • DOI
    10.1109/CPEM.2012.6250725
  • Filename
    6250725