Abstract :
The crucial point of absolute measurements of radiant power in the Terahertz spectral range is traceability to the International System of Units. PTB in Germany offers a reliable calibration of THz detectors as part of its official services as the only National Metrology Institute in the world. A dedicated calibration facility is based on a far-infrared molecular gas laser and a reference detector. Its power responsivity at 2.52 THz was originally measured in a pilot experiment using a cryogenic substitution radiometer. Actually the responsivity scale is expanded to the wider frequency range from 1 THz to 5 THz.
Keywords :
calibration; gas lasers; radiometers; terahertz wave detectors; units (measurement); Germany; International System of Units; PTB; THz detectors; cryogenic substitution radiometer; far-infrared molecular gas laser; frequency 1 THz to 5 THz; pilot experiment; radiant power measurements; reference detector; terahertz detectors; traceable calibration; Calibration; Detectors; Gas lasers; Optical detectors; Power measurement; Radiometry; Standards; Electromagnetic radiation; Terahertz detectors; gas lasers; laser stability; measurement uncertainty; optical detectors; power measurement;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements (CPEM), 2012 Conference on
Conference_Location :
Washington, DC
ISSN :
0589-1485
Print_ISBN :
978-1-4673-0439-9
Type :
conf
DOI :
10.1109/CPEM.2012.6250899
Filename :
6250899
Link To Document :
بازگشت