DocumentCode
2748856
Title
System for resistance unit transfer from 10 kΩ to 100 TΩ with new Hamon transfers
Author
Lisowski, Michal ; Krawczyk, Krystian
Author_Institution
Wroclaw Univ. of Technol., Wroclaw, Poland
fYear
2012
fDate
1-6 July 2012
Firstpage
382
Lastpage
383
Abstract
This paper describes realization in Poland of new system for resistance unit transfer from 10 kΩ to 100 TΩ, based on Hamon transfers. The new design of those transfers and their properties has been presented. Currently this system is being tested in Central Office of Measures in Poland.
Keywords
electric resistance measurement; Central Office of Measures; Hamon transfers; Poland; resistance 10 kohm to 100 Tohm; resistance unit transfer system; Connectors; Electrical resistance measurement; Insulation; Measurement uncertainty; Resistance; Resistors; Standards; Hamon transfers; measurement; measurement standards; measurement techniques; measurement uncertainty; precision measurements; uncertainty;
fLanguage
English
Publisher
ieee
Conference_Titel
Precision Electromagnetic Measurements (CPEM), 2012 Conference on
Conference_Location
Washington, DC
ISSN
0589-1485
Print_ISBN
978-1-4673-0439-9
Type
conf
DOI
10.1109/CPEM.2012.6250963