Title :
Dynamic calibration of bridge amplifiers used for periodical force measurement
Author :
Schlegel, C. ; Kieckenap, G. ; Kumme, R.
Author_Institution :
Phys.-Tech. Bundesanstalt, Braunschweig, Germany
Abstract :
This contribution describes how bridge amplifiers used in dynamic force measurement can be dynamically calibrated. A special analogue device, called “bridge standard”, can be used to simulate a bridge detuning in the mV range as, like it occurs in strain gauge force transducers [1]. The frequency response of the amplifier has to be corrected for the sensitivity of the force transducer. The principle of the bridge standard as well as measurements of different amplifiers will be presented. In addition, the influence of the amplifier frequency response of the parameter identification of the force transducer will be discussed.
Keywords :
bridge instruments; calibration; force measurement; force sensors; instrumentation amplifiers; amplifier frequency response; analogue device; bridge amplifier dynamic calibration; bridge standard; dynamic force measurement; force transducer parameter identification; periodical force measurement; strain gauge force transducers; Bridge circuits; Calibration; Dynamics; Force; Force measurement; Frequency response; Transducers;
Conference_Titel :
Precision Electromagnetic Measurements (CPEM), 2012 Conference on
Conference_Location :
Washington, DC
Print_ISBN :
978-1-4673-0439-9
DOI :
10.1109/CPEM.2012.6251057