• DocumentCode
    2750499
  • Title

    Dynamic calibration of bridge amplifiers used for periodical force measurement

  • Author

    Schlegel, C. ; Kieckenap, G. ; Kumme, R.

  • Author_Institution
    Phys.-Tech. Bundesanstalt, Braunschweig, Germany
  • fYear
    2012
  • fDate
    1-6 July 2012
  • Firstpage
    570
  • Lastpage
    571
  • Abstract
    This contribution describes how bridge amplifiers used in dynamic force measurement can be dynamically calibrated. A special analogue device, called “bridge standard”, can be used to simulate a bridge detuning in the mV range as, like it occurs in strain gauge force transducers [1]. The frequency response of the amplifier has to be corrected for the sensitivity of the force transducer. The principle of the bridge standard as well as measurements of different amplifiers will be presented. In addition, the influence of the amplifier frequency response of the parameter identification of the force transducer will be discussed.
  • Keywords
    bridge instruments; calibration; force measurement; force sensors; instrumentation amplifiers; amplifier frequency response; analogue device; bridge amplifier dynamic calibration; bridge standard; dynamic force measurement; force transducer parameter identification; periodical force measurement; strain gauge force transducers; Bridge circuits; Calibration; Dynamics; Force; Force measurement; Frequency response; Transducers;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements (CPEM), 2012 Conference on
  • Conference_Location
    Washington, DC
  • ISSN
    0589-1485
  • Print_ISBN
    978-1-4673-0439-9
  • Type

    conf

  • DOI
    10.1109/CPEM.2012.6251057
  • Filename
    6251057