DocumentCode
2750499
Title
Dynamic calibration of bridge amplifiers used for periodical force measurement
Author
Schlegel, C. ; Kieckenap, G. ; Kumme, R.
Author_Institution
Phys.-Tech. Bundesanstalt, Braunschweig, Germany
fYear
2012
fDate
1-6 July 2012
Firstpage
570
Lastpage
571
Abstract
This contribution describes how bridge amplifiers used in dynamic force measurement can be dynamically calibrated. A special analogue device, called “bridge standard”, can be used to simulate a bridge detuning in the mV range as, like it occurs in strain gauge force transducers [1]. The frequency response of the amplifier has to be corrected for the sensitivity of the force transducer. The principle of the bridge standard as well as measurements of different amplifiers will be presented. In addition, the influence of the amplifier frequency response of the parameter identification of the force transducer will be discussed.
Keywords
bridge instruments; calibration; force measurement; force sensors; instrumentation amplifiers; amplifier frequency response; analogue device; bridge amplifier dynamic calibration; bridge standard; dynamic force measurement; force transducer parameter identification; periodical force measurement; strain gauge force transducers; Bridge circuits; Calibration; Dynamics; Force; Force measurement; Frequency response; Transducers;
fLanguage
English
Publisher
ieee
Conference_Titel
Precision Electromagnetic Measurements (CPEM), 2012 Conference on
Conference_Location
Washington, DC
ISSN
0589-1485
Print_ISBN
978-1-4673-0439-9
Type
conf
DOI
10.1109/CPEM.2012.6251057
Filename
6251057
Link To Document