DocumentCode :
2752052
Title :
Development of a direct comparison system for radiofrequency power calibration transfer from 10 MHz to 18 GHz
Author :
Botello-Perez, Mariano ; Garcia-Ruiz, Israel
Author_Institution :
Centro Nac. de Metrol. CENAM, Mexico City, Mexico
fYear :
2012
fDate :
1-6 July 2012
Firstpage :
730
Lastpage :
731
Abstract :
This paper describes the development of a 10 MHz to 18 GHz radiofrequency power calibration transfer system based on the direct comparison technique using a resistive power splitter. This system is capable of measuring thermistor and thermoelectric based power detectors. The measurement uncertainties are thoroughly evaluated for the complex quantities in terms of magnitude and phase by following the internationally recommended guidelines. The uncertainty components for the system will be discussed and some measurement results for a power detector calibration will be presented.
Keywords :
calibration; measurement uncertainty; microwave measurement; power dividers; power measurement; thermistors; transfer standards; complex quantity; direct comparison system; frequency 10 MHz to 18 GHz; internationally recommended guidelines; measurement uncertainty; power detector calibration; radiofrequency power calibration transfer; resistive power splitter; thermistor based power detectors; thermoelectric based power detectors; uncertainty components; Calibration; Detectors; Measurement uncertainty; Power measurement; Radio frequency; Standards; Uncertainty; Microwave measurements; measurement uncertainty; power measurement; thermistor devices; transmission line measurements; uncertainty;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements (CPEM), 2012 Conference on
Conference_Location :
Washington, DC
ISSN :
0589-1485
Print_ISBN :
978-1-4673-0439-9
Type :
conf
DOI :
10.1109/CPEM.2012.6251137
Filename :
6251137
Link To Document :
بازگشت