DocumentCode :
2753255
Title :
Risk assessment for implanted cardiac defibrillator workers occupationally exposed to electromagnetic fields
Author :
Magne, Isabelle ; Souques, Martine
Author_Institution :
EDF R&D, Moret-sur-Loing, France
fYear :
2012
fDate :
17-21 Sept. 2012
Firstpage :
1
Lastpage :
5
Abstract :
The functioning of cardiac implants such as implanted cardiac defibrillator (ICD) may be perturbed by the exposure of the implanted person to electromagnetic field at workplace. We propose a method in order to help the occupational physician to take a decision when assessing the fitness for work in such a case, which is based on in situ assessment studies. We give here examples of exposure to 50 Hz magnetic field and to 380-400 MHz electromagnetic field.
Keywords :
biomedical equipment; cardiology; defibrillators; electromagnetic fields; prosthetics; risk management; electromagnetic fields; frequency 380 MHz to 400 MHz; frequency 50 Hz; implanted cardiac defibrillator; in situ assessment; risk assessment; Electromagnetic fields; Employment; Implants; Magnetic fields; Medical services; Probes; Substations; cardiac implant; electromagnetic field; fitness for work; occupational exposure;