Title :
Falling risk evaluation in elderly using miniature gyroscope
Author :
Najafi, Bahareh ; Blanc, Y. ; Robert, Philippe
Author_Institution :
Metrol. Lab., Swiss Fed. Inst. of Technol., Lausanne
Abstract :
The time of sit-stand and stand-sit transition and its duration were measured using a miniature gyroscope on the chest and a portable recorder on the waist. Based on a simple model and wavelet analysis, three parameters related to postural transition, namely the average and standard deviation of transition duration and the occurrence of abnormal successive transition, were provided for evaluating the falling risk. The comparison between two groups of elderly subjects (i.e. fallers and non-fallers) shows that these parameters were significantly correlated to falling risk (p<0.02). The integration of the sensor and the recorder in a same module allows providing a system especially adapted for ambulatory instrumentation and tele-alarm systems
Keywords :
biomechanics; biomedical equipment; geriatrics; gyroscopes; health hazards; physiological models; wavelet transforms; abnormal successive transition; ambulatory instrumentation; chest; elderly subjects; fallers; falling risk evaluation; miniature gyroscope; nonfallers; portable recorder; postural transition; simple model; tele-alarm systems; transition duration; waist; wavelet analysis; Gyroscopes; Hospitals; Instruments; Laboratories; Legged locomotion; Performance evaluation; Risk analysis; Senior citizens; Testing; Wavelet analysis;
Conference_Titel :
Microtechnologies in Medicine and Biology, 1st Annual International, Conference On. 2000
Conference_Location :
Lyon
Print_ISBN :
0-7803-6603-4
DOI :
10.1109/MMB.2000.893846