Abstract :
It is shown, by simple physical argument, that measurements of intensity with a radiation pressure balance should not agree with those based on calorimetric techniques. The conclusion is ultimately a consequence of the circumstance that radiation pressure measurements relate to wave momentum, while calorimetric methods relate to wave energy. Measurements with some typical ultrasound fields are performed with a novel type of hydrophone, and these allow an estimate to be made of the magnitude of the discrepancy to be expected between the two types of output measurement in a typical case
Keywords :
acoustic waves; bioacoustics; hydrophone; output measurements; radiation pressure; ultrasound fields; wave energy; wave momentum; Acoustic measurements; Biomedical engineering; Dentistry; Educational institutions; Energy measurement; Force measurement; Hospitals; Physics; Pressure measurement; Ultrasonic variables measurement;