• DocumentCode
    2782124
  • Title

    Research on the Possibility for Reflexotherapy MW Irradiator to Increase MW Electromagnetic Oscillations Penetration Depth into Biological Tissue

  • Author

    Belavskaya, S.V. ; Koltkov, D.S. ; Lisitsyna, L.I.

  • Author_Institution
    Novosibirsk State Technical University, Novosibirsk, Russia
  • fYear
    2005
  • fDate
    13-15 Dec. 2005
  • Firstpage
    194
  • Lastpage
    201
  • Abstract
    The results of calculations of MW electromagnetic (EM) oscillations penetration depth into biological tissue (BT) under treatment by MW irradiator for reflexotherapy are discussed. The irradiator was made of the segment of coaxial waveguide. It is shown that in order to satisfy the reflexotherapy requirements on locality and MW EM oscillations penetration depth into BT it is necessary to use coaxial waveguide having inner electrode radius ∼0.5 mm and outer electrode radius of ∼1 - 1.5 mm.
  • Keywords
    Biological tissues; Biomedical electrodes; Coaxial components; Electromagnetic waveguides; Finite element methods; Magnetic analysis; Magnetic devices; Medical treatment; Software packages; Vibrations;