DocumentCode
2787522
Title
Development of medical telemetry system by spread spectrum communication
Author
Kyoso, Masaki ; Uchiyama, Akihiko
Author_Institution
Sch. of Sci. & Eng., Waseda Univ., Tokyo, Japan
Volume
1
fYear
1996
fDate
31 Oct-3 Nov 1996
Firstpage
301
Abstract
In order to transmit patients´ data with high reliability in hospital, we have developed a spread spectrum medical telemetry system. In this study, we showed characteristics of electromagnetic noise from an electric knife by experiments, and showed system performance of spread spectrum system by ECG transmission experiments. From these results, it was shown that the spread spectrum system had acceptable reliability under the interference from the electric knife
Keywords
biomedical telemetry; demodulation; electrocardiography; electromagnetic interference; medical signal processing; spread spectrum communication; ECG transmission; EMI; electric knife; electromagnetic noise; high reliability data transmission; medical telemetry system; spread spectrum communication; spread spectrum modulation; system performance; Coagulation; Electrocardiography; Electromagnetic interference; Frequency measurement; Hospitals; Noise level; Receiving antennas; Signal processing; Spread spectrum communication; Telemetry;