• DocumentCode
    279278
  • Title

    Traceable dielectric measurements by resonance methods in the frequency range 1-30 MHz

  • Author

    Hill, G.J.

  • Author_Institution
    ERA Technol. Ltd., Leatherhead, UK
  • fYear
    1992
  • fDate
    7-10 Sep 1992
  • Firstpage
    108
  • Lastpage
    110
  • Abstract
    It is shown that traceable Q measurements on a stable inductor can provide the basis for dielectric measurements in the frequency range 1-30 MHz. It is also possible to adapt computer controlled four terminal impedance measuring techniques by coupling them with a resonance adaptor and suitable test fixture, so that dielectric measurements may be made to ±2% in permittivity and ±5μ rad in dielectric loss
  • Keywords
    Q-factor measurement; computerised instrumentation; dielectric loss measurement; inductors; permittivity measurement; 1 to 30 MHz; computer controlled; dielectric loss; dielectric measurements; four terminal impedance measuring techniques; permittivity; resonance adaptor; resonance methods; stable inductor; test fixture; traceable Q measurements;
  • fLanguage
    English
  • Publisher
    iet
  • Conference_Titel
    Dielectric Materials, Measurements and Applications, 1992., Sixth International Conference on
  • Conference_Location
    Manchester
  • Print_ISBN
    0-85296-551-6
  • Type

    conf

  • Filename