• DocumentCode
    279680
  • Title

    High accuracy frequency response measurements of mm-wave photodiodes using a DFB heterodyne system with a novel detection scheme

  • Author

    Humphreys, D.A. ; Park, C.A.

  • Author_Institution
    NPL, Teddington, UK
  • fYear
    1990
  • fDate
    32892
  • Firstpage
    42522
  • Lastpage
    42526
  • Abstract
    If the optical frequencies of two lasers can be accurately controlled then the response of the photodetector can be determined by measuring the heterodyne RF power and the photocurrents. The distributed feedback laser (DFB) incorporates a grating within the laser cavity to reduce the linewidth and give single frequency operation. A DFB heterodyne system operating at 1.53 μm is described. The system uses a novel detection scheme to overcome problems caused by the laser linewidth. The errors introduced by electrical reflections between the photodiode and the measurement system are discussed. Measurements of two research photodiodes at frequencies of up to 40 GHz are presented. The 3 dB frequencies of the diodes are 30.8±0.5 GHz and 26.9±0.5 GHz
  • Keywords
    characteristics measurement; distributed feedback lasers; measurement errors; photodiodes; semiconductor junction lasers; 1.53 micron; 26.9 GHz; 30.8 GHz; 40 GHz; DFB heterodyne; distributed feedback laser; electrical reflections; errors; frequency response measurements; laser linewidth; mm-wave photodiodes;
  • fLanguage
    English
  • Publisher
    iet
  • Conference_Titel
    Optical Detectors, IEE Colloquium on
  • Conference_Location
    London
  • Type

    conf

  • Filename
    189685