DocumentCode
2798533
Title
Effect of nano stripe carbonized-polymer electrode on high S/N ratio in electrochemical detection
Author
Naka, Keisuke ; Hayashi, Hiroki ; Senda, Mitsugi ; Shiraishi, Haruki ; Konishi, Satoshi
Author_Institution
Ritsumeikan Univ., Kyoto
fYear
2007
fDate
21-25 Jan. 2007
Firstpage
195
Lastpage
198
Abstract
This paper presents a scaling effect of the nano stripe electrode made of carbonized-polymer on the S/N ratio in electrochemical detections. It is well known that microelectrodes show higher sensitivity and higher S/N ratio. Downsizing of electrode, however, causes a reduction in signal current. Co-author has proposed a novel theoretical concept of a stripe electrode to solve this trade-off. The S/N ratio increases with downsizing the width of the stripe electrode because the noise current exponentially decreases. The nano stripe electrode with 450 nm in width and 15 mum in the center-to-center separation of the electrode could achieve 17.15 times higher S/N ratio than that of a planar electrode having the same corresponding area size. Furthermore, comparisons of experimental and theoretical S/N ratios will be discussed.
Keywords
electrochemical electrodes; microelectrodes; micromechanical devices; nanotechnology; center-to-center separation; electrochemical detection; microelectrodes; nanostripe carbonized-polymer electrode; Chemical technology; Chemistry; Electrochemical processes; Electrodes; Equations; Microelectrodes; Micromechanical devices; Organic materials; Signal analysis; Signal to noise ratio;
fLanguage
English
Publisher
ieee
Conference_Titel
Micro Electro Mechanical Systems, 2007. MEMS. IEEE 20th International Conference on
Conference_Location
Hyogo
ISSN
1084-6999
Print_ISBN
978-1-4244-095-5
Electronic_ISBN
1084-6999
Type
conf
DOI
10.1109/MEMSYS.2007.4433053
Filename
4433053
Link To Document