• DocumentCode
    2798841
  • Title

    Portable rotary-drive-type biochemical-analysis system based on magnetic bead-droplet handling

  • Author

    Nagao, N. ; Shikida, M. ; Takayanagi, K. ; Honda, H. ; Okochi, M. ; Ito, H. ; Sato, K.

  • Author_Institution
    Nagoya Univ., Nagoya
  • fYear
    2007
  • fDate
    21-25 Jan. 2007
  • Firstpage
    457
  • Lastpage
    460
  • Abstract
    By applying a single rotary motion to our previously developed magnetic bead-droplet handling mechanism, we newly developed transportation and agitation operation mechanisms aiming at realizing a portable-sized biochemical analysis system. We used two permanent magnets for transporting beads and another magnet coupled with multi-layered flat coils for agitation. They are all mounted on a rotational table, and the total size of the system is 60 times45 times35 mm. We experimentally confirmed that the developed system can do all the beads´ handlings (transportation, extraction, fusion, and agitation) by the rotational motion only.
  • Keywords
    biochemistry; chemical analysis; coils; micromechanical devices; agitation operation mechanisms; magnetic bead-droplet handling; micrototal analysis system; mu-TAS; multilayered flat coils; permanent magnets; portable rotary-drive-type biochemical-analysis system; rotational motion; rotational table; size 35 mm; size 45 mm; size 60 mm; Biochemical analysis; Coils; Couplings; Magnetic separation; Micromechanical devices; Motion analysis; Permanent magnets; Petroleum; Pumps; Transportation;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Micro Electro Mechanical Systems, 2007. MEMS. IEEE 20th International Conference on
  • Conference_Location
    Hyogo
  • ISSN
    1084-6999
  • Print_ISBN
    978-1-4244-095-5
  • Electronic_ISBN
    1084-6999
  • Type

    conf

  • DOI
    10.1109/MEMSYS.2007.4433070
  • Filename
    4433070