Title :
First results with high-resolution PET detector modules using wavelength-shifting fibers
Author :
Worstell, W. ; Johnson, O. ; Kudrolli, H. ; Zavarzin, V.
Author_Institution :
Dept. of Phys., Boston Univ., MA, USA
Abstract :
The authors have constructed and tested prototype PET detector modules incorporating wavelength-shifting fiber ribbons alternating with thin CsI(Na) scintillator plates in multilayer stacks. Modules with cross-sectional area of 11 cm×11 cm and with 9 mm of total crystal thickness have been tested. A reconstructed spatial resolution of 3 mm FWHM has been obtained for a line source, with an energy resolution of <20% and a time resolution of 25 ns FWHM. Energy measurements and coincidence timing are provided by 4 photomultipliers in an Anger array on one 11 cm×11 cm module face. Multiplexed fiber readout has been implemented with multianode photomultipliers coupled to several fiber ribbons. The methods and resolution achieved are readily extended to larger systems in a straightforward and cost-effective manner. Results of tests with several phantoms are presented
Keywords :
biomedical equipment; gamma-ray detection; photomultipliers; positron emission tomography; solid scintillation detectors; 25 ns; 9 mm; Anger array; CsI:Na; coincidence timing; crystal thickness; energy resolution; fiber ribbons; high-resolution PET detector modules; line source; medical diagnostic imaging; medical instrumentation; multianode photomultipliers; multilayer stacks; nuclear medicine; phantoms; thin CsI(Na) scintillator plates; wavelength-shifting fibers; Detectors; Energy measurement; Energy resolution; Nonhomogeneous media; Optical fiber testing; Photomultipliers; Positron emission tomography; Prototypes; Spatial resolution; Timing;
Conference_Titel :
Nuclear Science Symposium, 1997. IEEE
Conference_Location :
Albuquerque, NM
Print_ISBN :
0-7803-4258-5
DOI :
10.1109/NSSMIC.1997.670457