Abstract :
An ultrasonic imaging system is presented for diagnostics of post discotomic scarring. The system images parameters derived from a Maximum Likelihood Estimation of generalized K-distribution parameters. This approach is capable to discriminate between Rayleigh-, structural- and Pre-Rayleigh scattering. The system uses multidirectional echo data of the same cross-sectional plane, obtained by a 5 MHz sector scanner that is linearly moved over the human back. This leads to a resolution of the parameter images of approximately 1 mm×1 mm
Keywords :
biomedical ultrasonics; image resolution; maximum likelihood estimation; muscle; parameter estimation; surgery; ultrasonic scattering; 5 MHz; Rayleigh scattering; back surgery; cross-sectional plane; generalized K-distribution parameters; high resolution imaging; maximum likelihood estimation; medical diagnostic imaging; multidirectional echo data; muscle ultrasonic diagnosis; parameter images; post discotomic scarring diagnostics; preRayleigh scattering; structural scattering; Back; High-resolution imaging; Image resolution; Maximum likelihood estimation; Muscles; Rayleigh scattering; Speckle; Statistics; Surgery; Ultrasonic imaging;