    Clinical blood flow measurements using diffraction-grating transducers

  • Author

    Vilkomerson, D. ; Lyons, D. ; Chilipka, T. ; Delamere, M. ; Lopath, P. ; Palanchon, P. ; Shung, K.K.

  • Author_Institution
    EchoCath Inc., Princeton, NJ, USA
  • Volume
    2
  • fYear
    1998
  • fDate
    1998
  • Firstpage
    1501
  • Abstract
    Changing the phase or frequency of the signal driving a diffraction-grating transducer produces beams at known, multiple angles; these multiple angle beams can be used to make angle-independent Doppler measurements. This paper addresses three questions in developing a diffraction-grating instrument for measuring blood flow: how efficient and broadband can these transducers be, how close can such transducers be placed to blood vessels, and can the Doppler spectrum shape be used to determine the velocity profile
  • Keywords
    Doppler measurement; biomedical transducers; biomedical ultrasonics; blood flow measurement; finite element analysis; ultrasonic transducers; Doppler spectrum shape; FEA; angle-independent Doppler measurements; clinical blood flow measurements; diffraction-grating transducers; limited diffraction beam; multiple angle beams; transducer design; velocity profile; Blood flow; Blood vessels; Diffraction; Doppler measurements; Fluid flow measurement; Frequency; Instruments; Shape measurement; Transducers; Velocity measurement;