Abstract
We present a depth-dependent mismatched-filter pulse compression technique which uses the tissue attenuation as a parameter for appropriate compression filter design. The estimation of the attenuation solely from the reflected signal is performed using a least-squares approach in the frequency domain. Based on this estimation, a set of mismatched-filters is calculated to achieve the desired main-lobe width, SNR and sidelobe characteristics. Simulations and first experimental results are used to demonstrate the technique. The numerical effort necessary for the estimation algorithm is discussed
Keywords
acoustic signal processing; biomedical ultrasonics; filtering theory; image coding; image resolution; least squares approximations; medical image processing; pulse compression; ultrasonic absorption; SNR; chirp excitation; coded excitation imaging; compression filter design; depth-dependent mismatched filtering; digital beamformer; estimation algorithm; filter design parameter; frequency domain; image quality; least-squares approach; main-lobe width; medical US imaging; phantom simulations; pulse compression technique; range resolution; reflected signal; robust estimation; sidelobe characteristics; soft tissue; tissue attenuation; ultrasonic attenuation; Attenuation; Biomedical imaging; Chirp; Filtering; Filters; Frequency; Image coding; Pulse compression methods; Signal resolution; Ultrasonic imaging;
fLanguage
English
Publisher
ieee
Conference_Titel
Ultrasonics Symposium, 1998. Proceedings., 1998 IEEE
Conference_Location
Sendai
ISSN
1051-0117
Print_ISBN
0-7803-4095-7
Type
conf
DOI
10.1109/ULTSYM.1998.765260
Filename
765260
Link To Document