DocumentCode :
2808398
Title :
Multi-pulse ultrasound contrast imaging based on a decorrelation detection strategy
Author :
Frinking, Peter J A ; Céspedes, K.W. ; De Jong, Nico
Author_Institution :
Dept. of Cardiology & Exp. Echocardiography, Erasmus Univ., Rotterdam, Netherlands
Volume :
2
fYear :
1998
fDate :
1998
Firstpage :
1787
Abstract :
A novel technique for the detection of ultrasound contrast agents is described. The particular signature that occurs for encapsulated types of contrast agents, at high acoustical power settings, is associated with the release of free gas bubbles. Free gas bubbles show superior scattering characteristics compared to encapsulated gas bubbles. By correlating the response of two acoustic pulses before and after the release of gas bubbles, the ultrasound contrast agent can be exclusively detected
Keywords :
acoustic correlation; biomedical ultrasonics; bubbles; decorrelation; image enhancement; image resolution; medical image processing; medical signal detection; 2D image; acoustic pulses; colour-coded map; decorrelation detection strategy; echo signal change; encapsulated agents; high resolution imaging; multi-pulse ultrasound contrast imaging; release of free gas bubbles; scattering characteristics; ultrasound contrast agents; Acoustic imaging; Acoustic pulses; Acoustic scattering; Acoustic signal detection; Cardiology; Decorrelation; Frequency; High-resolution imaging; Image resolution; Ultrasonic imaging;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Ultrasonics Symposium, 1998. Proceedings., 1998 IEEE
Conference_Location :
Sendai
ISSN :
1051-0117
Print_ISBN :
0-7803-4095-7
Type :
conf
DOI :
10.1109/ULTSYM.1998.765296
Filename :
765296
Link To Document :
بازگشت