Title :
Single step algorithms for image reconstruction. II. Generalisation of the Sheffield filtered back projection algorithm. [Electric impedance tomography]
Author :
Avis, N J ; Barber, D.C.
Author_Institution :
Dept. of Comput. Sci., Hull Univ., UK
Abstract :
For pt.I see ibid. paper 4 (1992). The Sheffield system reconstructs images by a filtered back projection algorithm. To date the images have been produced from data originating from the adjacent drive configuration. However in principle the algorithm can be extended to reconstruct data originating from any bipolar drive configuration. Details of the various components of the image reconstruction process are given. Particular attention is paid to those areas where modifications are required to allow the technique to be extended to any dipolar drive configuration
Keywords :
computerised tomography; electric impedance imaging; image reconstruction; medical image processing; Sheffield filtered back projection algorithm; adjacent drive configuration; bipolar drive configuration; electric impedance tomography; image reconstruction;
Conference_Titel :
Electrical Impedance Tomography/Applied Potential Tomography, IEE Colloquium on
Conference_Location :
London