DocumentCode :
2812251
Title :
Multimodality image correlation quality assurance anthropomorphic phantom
Author :
Mutic, Sasa ; Bosch, Walter R. ; Drzymala, Robert E. ; Chao, K. S Clifford ; Low, Daniel A. ; Lewis, Jasons ; Cutler, P.D.
Author_Institution :
Radiat. Oncology Center, Washington Univ. Sch. of Med., St. Louis, MO, USA
Volume :
2
fYear :
2000
fDate :
2000
Firstpage :
1172
Abstract :
Multimodality images of many radiation therapy patients could be correlated/fused using software tools provided in treatment planning software. We have developed a phantom and a registration method for evaluation and verification of the fusion process and fusion software. Image registration was verified by contouring spheres and rods on the primary CT scan and observing contour locations on the second CT scan, MRI, or PET images which have been resampled in the plan of the primary CT scan . When properly registered, the contours of spheres and rods from the primary CT scan are expected to match images of these features in the other modalities. All contours were within 2 mm from their expected locations on the second CT scan, MRI, and PET scans. Proper registration was also verified by observing locations of fiducial markers, which were also within 2 mm from their expected locations. Again, the fiducial markers were outlined on the primary CT scan and their location observed on other scans
Keywords :
biomedical MRI; computerised tomography; image reconstruction; image registration; medical image processing; physiological models; positron emission tomography; radiation therapy; CT scans; MRI; PET images; contour locations; fiducial markers; fusion software; image reconstruction; image registration method; iterative method; multimodality image correlation; quality assurance anthropomorphic phantom; radiation therapy; software tools; treatment planning software; Anthropomorphism; Biomedical applications of radiation; Computed tomography; Image registration; Imaging phantoms; Magnetic resonance imaging; Medical treatment; Positron emission tomography; Quality assurance; Software tools;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society, 2000. Proceedings of the 22nd Annual International Conference of the IEEE
Conference_Location :
Chicago, IL
ISSN :
1094-687X
Print_ISBN :
0-7803-6465-1
Type :
conf
DOI :
10.1109/IEMBS.2000.897936
Filename :
897936
Link To Document :