DocumentCode :
2816666
Title :
Detailed description of FCs1: NRC’s cesium fountain primary standard
Author :
Marmet, L. ; Hoger, B. ; Dubé, P. ; Madej, A.A. ; Bernard, J.E.
Author_Institution :
Inst. for Nat. Meas. Stand., Nat. Res. Council, Ottawa, ON
fYear :
2008
fDate :
19-21 May 2008
Firstpage :
386
Lastpage :
391
Abstract :
The first cesium fountain primary standard built at the National Research Council Canada is described in details. Dimensional measurements and functions of components of the standard are given. We expect the standard to perform with a short term stability of 7times10-14 at one second and have a relative inaccuracy below the 1times10-15 level.
Keywords :
atomic clocks; caesium; frequency standards; Cs; NRC; National Research Council Canada; cesium fountain primary standard; dimensional measurements; frequency standards; Atomic beams; Coils; Councils; Fluorescence; Magnetic shielding; Measurement standards; Packaging; Pump lasers; Steel; Titanium;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Frequency Control Symposium, 2008 IEEE International
Conference_Location :
Honolulu, HI
ISSN :
1075-6787
Print_ISBN :
978-1-4244-1794-0
Electronic_ISBN :
1075-6787
Type :
conf
DOI :
10.1109/FREQ.2008.4623025
Filename :
4623025
Link To Document :