• DocumentCode
    2822286
  • Title

    A CMOS low noise amplifier for biomedical telemetry applications

  • Author

    Moreno, Robson Luiz ; Rodrigues, E.C.

  • Author_Institution
    Microelectron. Group, Univ. Fed. de Itajuba, Sao Paulo, Brazil
  • fYear
    2003
  • fDate
    5-7 May 2003
  • Firstpage
    95
  • Lastpage
    98
  • Abstract
    This paper describes a 200 MHz inductorless low-noise amplifier (LNA) fabricated in a standard digital 0.35-μm AMS CMOS process. The LNA has been measured to have a noise figure of 2.6 dB and a 24 dB gain at 200 MHz. The input IP3 is -12 dBm.
  • Keywords
    CMOS digital integrated circuits; biomedical telemetry; integrated circuit noise; radiofrequency amplifiers; semiconductor device noise; 0.35 micron; 2.6 dB; 200 MHz; 24 dB; CMOS low noise amplifier; LNA; biomedical telemetry applications; inductorless low-noise amplifier; noise figure; standard digital AMS CMOS process; Biomedical telemetry; CMOS process; CMOS technology; Capacitors; Circuits; Low-noise amplifiers; Noise figure; Operational amplifiers; Transconductance; Voltage;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Design, Test, Integration and Packaging of MEMS/MOEMS 2003. Symposium on
  • Print_ISBN
    0-7803-7066-X
  • Type

    conf

  • DOI
    10.1109/DTIP.2003.1287015
  • Filename