An LOR-based fully-3D PET image reconstruction using a blob-basis function
Author :
Hu, Z. ; Wang, W. ; Gualtieri, E.E. ; Hsieh, Y.L. ; Karp, J.S. ; Matej, S. ; Parma, M.J. ; Tung, C.H. ; Walsh, E.S. ; Werner, M. ; Gagnon, D.
Author_Institution :
Philips Med. Syst., Cleveland
Volume :
6
fYear :
2007
fDate :
Oct. 26 2007-Nov. 3 2007
Firstpage :
4415
Lastpage :
4418
Abstract :
Conventional reconstruction in Positron Emission Tomography (PET) imaging involves a line-of-response (LOR) preprocessing step where the raw LOR data are interpolated to evenly spaced sinogram data. The LOR-based reconstruction eliminates this interpolation step and thus gives rise to better spatial resolution and image quality. In the Philips PET/CT product, Gemini GXL, this approach is combined with a blob basis function that leads not only to substantial suppression of the image noise but also to preservation of the resolution. When projecting along the raw LORs, however, the computational advantage associated with projecting an evenly spaced sinogram is lost. In addition, using blobs to represent an object results in more image elements to trace in the projection because an LOR intersects more blobs than voxels for equivalent image quality. Therefore, the combined use of LOR-based reconstruction and a blob basis function requires significantly more computation time and represents a reconstruction performance challenge. In the Gemini GXL software we have used a system-matrix lookup table. Both multiplicative and additive corrections are modeled in the system matrix but not included in the lookup table. By making use of the scanner symmetry, and, more importantly, by aligning the blob matrix with the axial crystal rings, the lookup table is reduced in size by a factor of more than 100. The reconstruction performance is optimized by continuous memory access and block looping techniques in a hybrid-projection method. Compared to a calculate-on-the-fly approach, it is ~3 times faster on a Xeon 3.06 GHz dual-processor computer, which allows GXL to achieve excellent clinical performance.
Keywords :
image denoising; image reconstruction; image resolution; interpolation; medical image processing; positron emission tomography; table lookup; 3D PET image reconstruction; Gemini GXL software; blob-basis function; block looping technique; dual-processor computer, which; frequency 3.06 GHz; image noise suppression; image quality; line-of-response preprocessing; positron emission tomography; sinogram data; spatial resolution; system-matrix lookup table; Computed tomography; Image quality; Image reconstruction; Image resolution; Interpolation; Nuclear and plasma sciences; Optimization methods; Positron emission tomography; Spatial resolution; Table lookup;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record, 2007. NSS '07. IEEE
Conference_Location :
Honolulu, HI
ISSN :
1095-7863
Print_ISBN :
978-1-4244-0922-8
Electronic_ISBN :
1095-7863
Type :
conf
DOI :
10.1109/NSSMIC.2007.4437091
Filename :
4437091
Link To Document :
بازگشت