• DocumentCode
    2846666
  • Title

    Dissolution profiles of chloroquine tablets using a flow-through chloroquine sensor

  • Author

    Zain, Zainiharyati M. ; Saad, Bahruddin ; Rahman, Ismail Ab ; Mahsufi, Shariff

  • Author_Institution
    Dept. of Appl. Sci., Universiti Teknologi Mara, Penang, Malaysia
  • fYear
    2005
  • fDate
    5-7 Sept. 2005
  • Firstpage
    22
  • Lastpage
    25
  • Abstract
    A potentiometric polyvinyl chloride membrane comprising of bis(2-ethylhexyl)adipate as plasticizer and potassium tetrakis(4-chlorophenyl)borate was developed as a chloroquine sensor. The sensor was later applied as a detector in a flow injection analysis set up where it was not only successfully used in the determination of active ingredient in pharmaceutical preparations but also in establishing the dissolution profiles for chloroquine tablets. The flow through chloroquine sensor offers high accuracy and excellent repeatability, with large sample throughput (90 mL min-1). Besides being simple, the technique is rapid, low cost and allows direct quantification of chloroquine (without sample pretreatment).
  • Keywords
    chemical sensors; dissolving; drugs; membranes; polymers; bis(2-ethylhexyl)adipate; chloroquine tablets; dissolution; flow injection analysis; flow-through chloroquine sensor; plasticizer; potassium tetrakis(4-chlorophenyl)borate; potentiometric polyvinyl chloride membrane; Automation; Biomembranes; Chemical sensors; Detectors; Drugs; Electrodes; Fabrication; Pharmaceuticals; Quality control; Testing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Sensors and the International Conference on new Techniques in Pharmaceutical and Biomedical Research, 2005 Asian Conference on
  • Print_ISBN
    0-7803-9370-8
  • Type

    conf

  • DOI
    10.1109/ASENSE.2005.1564497
  • Filename
    1564497