DocumentCode :
2853270
Title :
Functional testing of digital microfluidic biochips
Author :
Xu, Tao ; Chakrabarty, Krishnendu
Author_Institution :
Dept. of Electr. & Comput. Eng., Duke Univ., Durham, NC
fYear :
2007
fDate :
21-26 Oct. 2007
Firstpage :
1
Lastpage :
10
Abstract :
Dependability is an important attribute for microfluidic biochips that are used for safety-critical applications such as point-of-care health assessment, air-quality monitoring, and food-safety testing. Therefore, these devices must be adequately tested after manufacture and during bioassay operations. Known techniques for biochip testing are all function-oblivious, i.e., while they can detect and locate defect sites on a microfluidic array, they cannot be used to ensure correct operation of functional units. In this paper, we introduce the concept of functional testing of microfluidic biochips. We address fundamental biochip operations such as droplet dispensing, droplet transportation, mixing, splitting, and capacitive sensing. Long electrode actuation times are avoided to ensure that there is no electrode degradation during testing. We evaluate the proposed test methods using simulations as well as experiments for a fabricated biochip.
Keywords :
biomolecular electronics; microfluidics; testing; air-quality monitoring; capacitive sensing; dependability; digital microfluidic biochips; droplet dispensing; droplet transportation; food-safety testing; functional testing; mixing; point-of-care health assessment; safety-critical application; splitting; Application software; Biological system modeling; Circuit testing; Computerized monitoring; Degradation; Electrodes; Food manufacturing; Microfluidics; Proteins; Transportation;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Test Conference, 2007. ITC 2007. IEEE International
Conference_Location :
Santa Clara, CA
ISSN :
1089-3539
Print_ISBN :
978-1-4244-1127-6
Electronic_ISBN :
1089-3539
Type :
conf
DOI :
10.1109/TEST.2007.4437614