Direct calibration of a reference chamber against the low air kerma rate primary standard at Ir-192 energy
Author :
Rajan, K. N Govinda ; Palaniselvan, N. ; Bhatt, B.C. ; Vijayam, M. ; Patki, V.S. ; Vinatha ; Jadhavgaonkar ; Pendse, A.M. ; Kannan, V.
Author_Institution :
Bhabha Atomic Res. Centre, Bombay, India
Volume :
4
fYear :
2000
fDate :
2000
Firstpage :
2867
Abstract :
Using a 60 cc spherical graphite ionization chamber as the Primary Standard the HDR Ir-192 source was standardized at the hospital site in a minimum scatter geometry, in units of Air Kerma Strength (AKS). A 400 cc bakelite reference chamber was calibrated against the Primary Standard. It was seen that the Primary Standard and the Reference chamber, both being of low Z, showed roughly the same scatter response and yielded the same calibration factor for the 400 cc reference chamber, with or without scatter. The scatter correction, however, is required for the determination of the Air kerma strength. The theoretical Monte Carlo estimates of the scatter corrections, obtained by simulating the source and the room geometry, showed excellent agreement with the experimental values. This reference chamber can now be used at other hospital sites for determining the AKS of the source, by applying proper scatter corrections. However, any likelihood of change in the reference chamber calibration would necessitate the transport of the Primary Standard to the hospital site for recalibration. Frequent transport of the Primary Standard can lead to changes in its response due to movement or other extraneous causes. The calibration of the Reference chamber against the Primary Standard at the Standards Lab., for an industrial type Ir-192 source maintained at the lab., showed excellent agreement with the hospital calibration, making it possible to check the reference chamber calibration at the lab. itself. Further work is in progress
Keywords :
Monte Carlo methods; calibration; ionisation chambers; iridium; radiation therapy; radioactive sources; radioisotopes; 400 cc bakelite reference chamber; 60 cc spherical graphite ionization chamber; HDR Ir-192 source; Ir; Ir-192 energy; Primary Standard; calibration factor; direct calibration; high dose rate brachytherapy; hospital site; low air kerma rate primary standard; minimum scatter geometry; reference chamber; scatter corrections; theoretical Monte Carlo estimates; therapeutic nuclear medicine; Attenuation; Brachytherapy; Calibration; Dosimetry; Geometry; Hospitals; Ionization chambers; Laboratories; Measurement standards; Scattering;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society, 2000. Proceedings of the 22nd Annual International Conference of the IEEE
Conference_Location :
Chicago, IL
ISSN :
1094-687X
Print_ISBN :
0-7803-6465-1
Type :
conf
DOI :
10.1109/IEMBS.2000.901471
Filename :
901471
Link To Document :
بازگشت