EMC and Doppler ultrasound equipment
Author :
Follett, D.H.
Author_Institution :
Dept. of Med. Phys., Bristol Gen. Hospital, UK
fYear :
1993
fDate :
34067
Firstpage :
42644
Lastpage :
42647
Abstract :
The work described has allowed proposals to be made on a reasonably sound basis for testing susceptibility to interference entering via the two routes found to be most important in practice. The tests can be carried out with only basic laboratory equipment comprising a simple signal generator and an oscilloscope. The main problem remaining is that due to the vagaries of the signal injection method the actual test levels are uncertain and match the required profile only over a restricted frequency range
Keywords :
biomedical equipment; biomedical ultrasonics; electromagnetic compatibility; electromagnetic interference; Doppler ultrasound equipment; EMI paths; electromagnetic compatibility; susceptibility to interference;