DocumentCode :
2867110
Title :
Computational and experimental studies of orthopedic implants heating under MRI RF coils
Author :
Liu, Yan ; Chen, Ji ; Shellock, Frank ; Kainz, Wolfgang
Author_Institution :
Department of Electrical and Computer Engineering, University of Houston, TX 77204, USA
fYear :
2012
fDate :
17-22 June 2012
Firstpage :
1
Lastpage :
3
Abstract :
The heating of orthopedic implants under MRI RF fields were investigated 1.5T and 3T systems. Modeling and experiments were performed on an orthopedic device at different sizes inside an ASTM phantom. It is observed that the induced energy deposition near the device is almost linearly related to the dimension of the orthopedic implants when the device is less than 10 cm in length. Higher temperature rises were found in 1.5T for both computational and experimental studies.
Keywords :
Coils; Conductors; Heating; Magnetic field measurement; Magnetic resonance; Radio frequency; MRI Safety; electromagnetic modeling; heating; medical implants;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Microwave Symposium Digest (MTT), 2012 IEEE MTT-S International
Conference_Location :
Montreal, QC, Canada
ISSN :
0149-645X
Print_ISBN :
978-1-4673-1085-7
Electronic_ISBN :
0149-645X
Type :
conf
DOI :
10.1109/MWSYM.2012.6259617
Filename :
6259617
Link To Document :