DocumentCode :
2872956
Title :
The optical blood test device with the micro Fabry-Perrot interferometer
Author :
DOHI, Tetsuji ; Matsumoto, Kiyoshi ; Shimoyama, Isao
Author_Institution :
Dept. of Mech.-Inf., Tokyo Univ., Japan
fYear :
2004
fDate :
2004
Firstpage :
403
Lastpage :
406
Abstract :
This paper reports on an optical blood test device using a micro Fabry-Perot interferometer (micro FPI) implanted in a body. Since we intend to measure many kinds of information from blood by using only one device, we have developed a micro FPI. A prototype of the micro FPI was fabricated beforehand to check the characteristics of the micro FPI. The prototype can be used in wide wavelength ranging from visible light to near-infrared. It transmits the light from 30 to 80% in the range. The light transmittances of arterial blood and venous blood were measured by the prototype. Based on the prototype, a micro FPI was fabricated. The characteristics of the micro FPI are similar to the prototype. It turns out that this micro FPI has sufficient characteristics to measure the oxygen concentration.
Keywords :
bio-optics; biomedical equipment; biosensors; blood; light interferometry; microsensors; arterial blood; micro Fabry-Perrot interferometer; optical blood test device; oxygen concentration; venous blood; visible light; wavelength; Biomedical optical imaging; Biosensors; Blood; Optical devices; Optical filters; Optical interferometry; Optical sensors; Prototypes; Testing; Wavelength measurement;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Micro Electro Mechanical Systems, 2004. 17th IEEE International Conference on. (MEMS)
Print_ISBN :
0-7803-8265-X
Type :
conf
DOI :
10.1109/MEMS.2004.1290607
Filename :
1290607
Link To Document :
بازگشت