• DocumentCode
    2875748
  • Title

    Advanced X-ray Detectors for Volumetric CT

  • Author

    Deych, Ruvin ; Dolazza, Enrico

  • Author_Institution
    Analogic Corp., Peabody
  • fYear
    2007
  • fDate
    6-8 Feb. 2007
  • Firstpage
    1
  • Lastpage
    6
  • Abstract
    The rapid evolution of X-ray computed tomography (CT) in the past decade, from single slice axial scanning to 256-slice helical imaging, was driven by remarkable progress in X-ray generation and detection and by continuously lower cost of high speed computation. In this paper, we present the characteristics of the X-ray detectors required for today´s multi-slice CT scanners suitable for cardiac applications. In particular, the parametric requirements for isotropic, solid-state, scintillator based, X-ray detectors and their associated electronics used in volumetric CT imaging are described.
  • Keywords
    X-ray detection; cardiology; computerised tomography; X-ray detectors; X-ray generation; cardiac applications; helical imaging; high speed computation; slice axial scanning; volumetric computed tomography imaging; Abdomen; Computed tomography; Consumer electronics; Image reconstruction; Optical imaging; Pain; USA Councils; X-ray detection; X-ray detectors; X-ray imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Sensors Applications Symposium, 2007. SAS '07. IEEE
  • Conference_Location
    San Diego, CA
  • Print_ISBN
    1-4244-0678-1
  • Electronic_ISBN
    1-4244-0678-1
  • Type

    conf

  • DOI
    10.1109/SAS.2007.374374
  • Filename
    4248496