Title :
An evaluation of XCTD performance with design improvements
Author :
Elgin, Randall H.
Author_Institution :
Ocean Prod. Div., Sippican Inc., Marion, MA, USA
Abstract :
In the last quarter of 1993 Sippican underwent an extensive evaluation of the XCTD/MK12 System. This included a review of the existing system, 7 local tow tests with a reference CTD, 3 sea tests off Bermuda and in-house tests based on regular sampling of the production line. The authors confirmed some of the performance difficulties encountered by users-notably excessive noise and offsets of temperature and/or conductivity at 900 m. Several modifications have been made to improve the overall system performance. Testing of the improved system was done on 18 probes tow tested with the reference CTD and 48 probes dropped off Bermuda. The results showed mean temperature and conductivity accuracy errors of -0.013 ° C and -0.003 mS/cm. The range of accuracy errors was -0.03 to 0.01 °C and -0.03 to 0.02 mS/cm, with standard deviations of 0.012 and 0.013 respectively
Keywords :
electric sensing devices; electrical conductivity measurement; oceanographic equipment; terrestrial electricity; terrestrial heat; testing; thermometers; Sippican; XCTD performance; XCTD/MK12 System; accuracy errors; conductivity; design improvements; excessive noise; in-house tests; performance; temperature; tow tests; Conductivity measurement; Grounding; Noise measurement; Numerical models; Ocean temperature; Probes; Sampling methods; System performance; System testing; Temperature measurement;
Conference_Titel :
OCEANS '94. 'Oceans Engineering for Today's Technology and Tomorrow's Preservation.' Proceedings
Conference_Location :
Brest
Print_ISBN :
0-7803-2056-5
DOI :
10.1109/OCEANS.1994.363868