Abstract :
A precise rotating coil measurement system (RCS) is developed to characterize the magnetic field quality of the quadrupole (QM) and sextupole (SM) magnets in the Taiwan photon source (TPS). A measurement bench is designed to install the magnets easily and mount the rotating coil unit with high reproducibility. The fiberglass reinforced epoxy (FRP) measurement unit (F-unit) exhibits a large sag and mechanical error while it is 880 mm long. Therefore, a new graphite measurement unit (G-unit) with a printed circuit coil is adopted to reduce these errors. The rotating coil design and testing using a QM are also described.
Keywords :
accelerator magnets; coils; light sources; magnetic fields; synchrotron radiation; synchrotrons; RCS; TPS; Taiwan photon source; fiberglass reinforced epoxy; graphite measurement unit; magnetic field; printed circuit coil; quadrupole magnets; rotating coil measurement system; sextupole magnets; synchrotron radiation; Coils; Fiber reinforced plastics; Magnetic field measurement; Magnets; Measurement units; Optical fiber testing; Printed circuits; Reproducibility of results; Rotation measurement; Samarium;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Particle Accelerator Conference, 2007. PAC. IEEE
Conference_Location :
Albuquerque, NM
Print_ISBN :
978-1-4244-0916-7
Type :
conf
DOI :
10.1109/PAC.2007.4440220
Filename :
4440220
Link To Document :
بازگشت