• DocumentCode
    2882934
  • Title

    A moving average based filtering system with its application to real-time QRS detection

  • Author

    Chen, HC ; Chen, SW

  • Author_Institution
    Dept. of Electr. Eng., Chang Gung Univ., Taiwan
  • fYear
    2003
  • fDate
    21-24 Sept. 2003
  • Firstpage
    585
  • Lastpage
    588
  • Abstract
    This paper presents a novel real-time QRS detection algorithm designed based on a simple moving average filter. The proposed algorithm demands no redundant preprocessing step, thus allowing a simple architecture for its implementation as well as low computational cost. Algorithm performance was validated against a subset of the MIT-BIH arrhythmia database. Consequently, numerical results showed that the proposed algorithm correctly detected over 99.5% of the QRS complexes from the standard ECG database, implying it may be considered as a simple and reliable candidate of QRS detection algorithms.
  • Keywords
    electrocardiography; high-pass filters; low-pass filters; medical signal detection; moving average processes; nonlinear filters; ECG database; MIT-BIH arrhythmia database; moving average based filtering system; real-time QRS detection; Algorithm design and analysis; Computational efficiency; Databases; Delay; Detection algorithms; Electrocardiography; Filtering; Finite impulse response filter; Nonlinear filters; Real time systems;