Title :
The Prediction of the E.M.F. Stability of the High Precision Saturated Cadmium Sulfate Standard Cells
Author :
Qu Jin ; Zhu Xun-Zhi
Author_Institution :
Yangzhou Teacher´´s College
Abstract :
This paper briefly analyses the various factors which may cause the cell´s emf increasing or decreasing. The different method to predict the emf stability of 1-2 ppm high precision standard cells has been described here. Experimental results have shown that among these methods the emf hysteresis effect test method recommended in GB3929-83 is most valuable.
Conference_Titel :
Precision Electromagnetic Measurements, 1988. CPEM 88 Digest. 1988 Conference on
Conference_Location :
Ibaraki, Japan
DOI :
10.1109/CPEM.1988.671234