DocumentCode
2901590
Title
Primary thermometry by the measurement of shot noise in a tunnel junction
Author
Song, Woon ; Park, Jung Hwan ; Rehman, Mushtaq ; Chong, Yonuk
Author_Institution
Center for Nano & Quantum Sci., Korea Res. Inst. of Stand. & Sci., Daejeon, South Korea
fYear
2011
fDate
12-16 June 2011
Firstpage
88
Lastpage
89
Abstract
We fabricated an Al-AlOx-Al tunnel junction by the double-angle evaporation technique as a source of shot noise. A precision RF-power-measurement setup in the frequency range near 1 GHz was installed in a He-3 refrigerator. We measured temperature from 0.3 K to room temperature using shot noise thermometer and compared the temperature with that of conventional temperature sensors.
Keywords
alumina; noise measurement; power measurement; refrigerators; shot noise; thermometers; Al-AlOx-Al; He-3 refrigerators; RF-power-measurement setup; double-angle evaporation technique; primary thermometry; shot noise thermometer; temperature 0.3 K; temperature sensors; tunnel junction; Fitting; Junctions; Noise; Noise measurement; Temperature measurement; Temperature sensors; Voltage measurement; Cryogenics; Primary Thermometry; Shot noise; Tunneling;
fLanguage
English
Publisher
ieee
Conference_Titel
Noise and Fluctuations (ICNF), 2011 21st International Conference on
Conference_Location
Toronto, ON
Print_ISBN
978-1-4577-0189-4
Type
conf
DOI
10.1109/ICNF.2011.5994392
Filename
5994392
Link To Document