DocumentCode
2904074
Title
Accurate RF Voltage Measurements Using a Sampling Voltage Tracker
Author
Souders, T.M. ; Hetrick, P.S.
Author_Institution
National Bureau of Standards
fYear
1988
fDate
7-10 June 1988
Firstpage
270
Lastpage
271
Abstract
The rf voltage measurement capability of an equivalent time sampling system is described. The frequency range investigated is 1 to 100 NHz. Over this range, the measured errors, determined by ac/dc thermal transfer, are within the stated uncertainties presently provided by NBS for thermal converter calibrations. The system offers several advantages over conventional thermal transfer techniques: ac/dc transfers are not required, loading and transmission line problems are reduced, and direct measurement of voltages from 2 V to as low as 10 mV are possible.
fLanguage
English
Publisher
ieee
Conference_Titel
Precision Electromagnetic Measurements, 1988. CPEM 88 Digest. 1988 Conference on
Conference_Location
Ibaraki, Japan
Type
conf
DOI
10.1109/CPEM.1988.671291
Filename
671291
Link To Document