  • Abstract
    In a study performed to test the effects of proton and gamma radiation on different temperature sensors, Pt RTDs and thermocouples show a negligible sensitivity. On the other hand, IC-based temperature sensors modify significantly their behavior. AD590 sensors exhibit a moderate degradation, depending on the encapsulation. And TMP36 sensors experiment a big degradation amounting up to a change in the readout signal about 20°C for a total ionizing dose of 40 krad.
  • Keywords
    encapsulation; gamma-ray effects; proton effects; readout electronics; temperature sensors; thermocouples; AD590 sensor; IC-based temperature sensor; RTD; TMP36 sensor experiment; gamma radiation; proton radiation; readout signal; thermocouple; total ionizing dose; Protons; Radiation effects; Sensitivity; Temperature distribution; Temperature measurement; Temperature sensors; Temperature sensors; bipolar linear circuits; radiation effects; total ionizing dose;