Title :
Clinical evaluation of single-photon attenuation correction for 3D whole-body PET
Author :
Watson, C.C. ; Schaefer, A. ; Luk, W.K. ; Kirsch, C.-M.
Author_Institution :
CTI PET Syst. Inc., Knoxville, TN, USA
Abstract :
Reports on the first clinical evaluation of the 137Cs single photon transmission measurement that has been implemented on the Siemens/CTI ECAT ART. Thorax phantom studies are performed to verify the accuracy of the transmission measurement. Pre-injection (with patient repositioning) and post-injection transmission scans and attenuation corrected emission images of 25 patients are compared with respect to estimated linear attenuation coefficients, specific activities and standardized uptake values in lesions. The authors find that both pre- and post-injection transmission provide accurate estimates of attenuation correction factors. Post-injection estimates for attenuation coefficients and specific activities in larger regional ROIs are nearly the same as pre-injection values, but exhibit 18% greater pixel standard deviation. Estimates of standardized uptake in smaller lesion ROIs are 13% higher with post-injection correction, but there is no difference in the ROI pixel variances
Keywords :
gamma-ray absorption; medical image processing; positron emission tomography; 3D whole-body PET; Cs; ROI pixel variances; attenuation coefficients; attenuation correction factors; clinical evaluation; lesion ROIs; medical diagnostic imaging; nuclear medicine; patient repositioning; pixel standard deviation; pre-injection; single-photon attenuation correction; specific activities; thorax phantom studies; transmission measurement accuracy verification; Attenuation measurement; Collimators; Imaging phantoms; Lesions; Noise measurement; Performance evaluation; Subspace constraints; Technological innovation; Thorax; Whole-body PET;
Conference_Titel :
Nuclear Science Symposium, 1998. Conference Record. 1998 IEEE
Conference_Location :
Toronto, Ont.
Print_ISBN :
0-7803-5021-9
DOI :
10.1109/NSSMIC.1998.773867