Measurement of Acoustic Output of Ultrasound Instrumentation for Interstitial Thermal Therapy
Author
Jarosz, Boguslaw J.
Author_Institution
Ottawa-Carleton Inst. for Phyics., Carleton Univ., Ottawa, ON
fYear
2007
fDate
1-3 May 2007
Firstpage
1
Lastpage
5
Abstract
A new ultrasound interstitial instrumentation is tested for its use as a thermal therapy device. The acoustic output of the device is measured to enable evaluation of the device performance in intended heating of abnormal tissues. We propose figure of merit for acoustic signal processing to evaluate device comparison. Our results show that the new applicator with compression type piezoelectric element has a promise as a therapeutic device. The shear type applicator seems to be impractical.
Keywords
acoustic signal processing; biomedical ultrasonics; biothermics; medical signal processing; patient treatment; piezoelectric devices; ultrasonic equipment; abnormal tissues; acoustic output measurement; acoustic signal processing; interstitial thermal therapy; piezoelectric element; therapeutic device; thermal therapy device; ultrasound instrumentation; ultrasound interstitial instrumentation; Acoustic devices; Acoustic measurements; Acoustic signal processing; Acoustic testing; Applicators; Heating; Instruments; Medical treatment; Ultrasonic imaging; Ultrasonic variables measurement; acoustic measurements; cancer treatment; medical instrumentation; thermal therapy; ultrasound instrumentation;
fLanguage
English
Publisher
ieee
Conference_Titel
Instrumentation and Measurement Technology Conference Proceedings, 2007. IMTC 2007. IEEE
Conference_Location
Warsaw
ISSN
1091-5281
Print_ISBN
1-4244-0588-2
Type
conf
DOI
10.1109/IMTC.2007.379049
Filename
4258338
Link To Document