Title :
Post-Correction of Under-Sampled Analog to Digital Converters
Author :
Björsell, Niclas ; Händel, Peter
Author_Institution :
Gavle Univ., Gavle
Abstract :
Applications with wide bandwidth and high center frequencies force the analog to digital converter (ADC) to be active in a working range with less dynamic performance in relation to lower frequency bands. However using under-sampling techniques in combination with post-correction methods enable a combination of high sampling rate, wide bandwidth and low distortion. In this paper the employed dynamic post-correction is a combination of look-up tables and model based correction. The results are mainly based on measurements on a 12-bit 210 MSPS ADC. The improvement in total harmonic distortion and spurious free dynamic range are acceptable over a wide frequency range and it is robust to variations in amplitude.
Keywords :
analogue-digital conversion; harmonic distortion; table lookup; analog-digital converters; dynamic post correction; look-up tables; lower frequency bands; post-correction methods; total harmonic distortion; under-sampling techniques; word length 12 bit; Analog-digital conversion; Bandwidth; Distortion measurement; Dynamic range; Frequency conversion; Image sampling; Instrumentation and measurement; Sampling methods; Table lookup; Total harmonic distortion; ADC; Analog to digital converters; Measurements; Post-correction; Test; Under-sampling;
Conference_Titel :
Instrumentation and Measurement Technology Conference Proceedings, 2007. IMTC 2007. IEEE
Conference_Location :
Warsaw
Print_ISBN :
1-4244-0588-2
DOI :
10.1109/IMTC.2007.379429