DocumentCode :
2915526
Title :
Transcranial magnetic stimulation in the presence of deep brain stimulation implants: Induced electrode currents
Author :
Deng, Zhi-De ; Lisanby, Sarah H. ; Peterchev, Angel V.
Author_Institution :
Dept. of Electr. Eng., Columbia Univ., New York, NY, USA
fYear :
2010
fDate :
Aug. 31 2010-Sept. 4 2010
Firstpage :
6821
Lastpage :
6824
Abstract :
The safety of transcranial magnetic stimulation (TMS) in patients with an implanted deep brain stimulation (DBS) systems has not been thoroughly investigated. One potential safety hazard is the induction of significant voltages in the subcutaneous leads in the scalp that could result in unintended electrical currents in the DBS electrode contacts. We measured ex-vivo the TMS-induced voltages and currents in DBS electrodes with the implantable pulse generator (IPG) set in various modes of operation. We show that voltages as high as 100 V resulting in currents as high as 83 mA can be induced in the DBS leads by a TMS pulse in all IPG modes. These currents are an order of magnitude higher than the normal DBS pulses, and could result in tissue damage. When the IPG is turned off, electrode currents flow only if the TMS-induced voltage exceeds 5 V.
Keywords :
bioelectric potentials; biological tissues; biomagnetism; biomedical electrodes; brain; prosthetics; pulse generators; IPG modes; TMS-induced voltages; deep brain stimulation implants; implantable pulse generator; induced electrode currents; scalp; subcutaneous leads; tissue damage; transcranial magnetic stimulation; Brain stimulation; Coils; Electrodes; Impedance; Lead; Satellite broadcasting; Voltage measurement; Brain; Deep Brain Stimulation; Electric Stimulation; Electrodes, Implanted; Humans; Transcranial Magnetic Stimulation;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society (EMBC), 2010 Annual International Conference of the IEEE
Conference_Location :
Buenos Aires
ISSN :
1557-170X
Print_ISBN :
978-1-4244-4123-5
Type :
conf
DOI :
10.1109/IEMBS.2010.5625958
Filename :
5625958
Link To Document :
بازگشت