DocumentCode :
2916917
Title :
Uncertainty in a Calibration of a Photodiode Type Power Meter Due to Spectral Width of a Light Source
Author :
Inoue, T. ; Yokoshima, I. ; Katada, F.
Author_Institution :
Electrotechnical Laboratory
fYear :
1988
fDate :
7-10 June 1988
Firstpage :
395
Lastpage :
396
Abstract :
An error due to a spectral width of a light source in a calibration of a photodiode type power meter has been studied. Three types of Si, Ge and InGaAs photodiode have been examined in the wavelength of 750~1700nm. A theoretical calculation has shown that more than 5% error arises on a Si or a Ge photodiode in a specific wavelength region. A measured difference between two photodiode type power meter agreed the theoretically estimated difference within 0.5%.
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements, 1988. CPEM 88 Digest. 1988 Conference on
Conference_Location :
Ibaraki, Japan
Type :
conf
DOI :
10.1109/CPEM.1988.671367
Filename :
671367
Link To Document :
بازگشت