• DocumentCode
    2932001
  • Title

    A complete calibration procedure for time domain network analyzers

  • Author

    Ferrari, P. ; Angenieux, G. ; Flechet, B.

  • Author_Institution
    LAHC, Univ. de Savoie, Le Bourget du Lac, France
  • fYear
    1992
  • fDate
    1-5 June 1992
  • Firstpage
    1451
  • Abstract
    A complete calibration procedure for a fast sampling oscilloscope has been developed. This calibration makes it possible to use time-domain-reflected, time-domain-transmitted (TDR-TDT) systems at a quantitative level as a time-domain network analysis (TDNA) system. The data acquisition and calibration programs are implemented on a personal computer. All measurements are achieved in the time domain; the errors are corrected in the frequency domain, after Fourier transform. The bandwidth of the system extends from DC to 20 GHz. A whole calibration is achieved in a time comparable to LRL calibration on a classical network analyzer.<>
  • Keywords
    calibration; microwave measurement; network analysers; time-domain reflectometry; 0 to 20 GHz; Fourier transform; calibration procedure; data acquisition; fast sampling oscilloscope; time domain network analyzers; time-domain network analysis; time-domain reflected time-domain transmitted systems; Calibration; Computer errors; Data acquisition; Error correction; Frequency measurement; Microcomputers; Oscilloscopes; Sampling methods; Time domain analysis; Time measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Microwave Symposium Digest, 1992., IEEE MTT-S International
  • Conference_Location
    Albuquerque, NM, USA
  • ISSN
    0149-645X
  • Print_ISBN
    0-7803-0611-2
  • Type

    conf

  • DOI
    10.1109/MWSYM.1992.188283
  • Filename
    188283